Building the force-matching rig

What the device page records is where the instrument ended up. This is the part that record leaves out: what had to be decided, and what I got wrong on the way.

A force-matching rig on a desk. A metal power supply unit sits inside a clear acrylic enclosure on a transparent base plate, wired in red and yellow to a small controller board behind it. A black cylindrical motor is held in a white 3D-printed mount, with a curved metal lever arm rising from it. A slider potentiometer with a yellow thumb tab lies loose on the plate. Mains cables run off the desk, and books and a block of grey packing foam sit behind the rig.

The rig on the bench. From the left: the power supply in its enclosure, the drive behind it, the motor in its printed mount with the lever arm above, and the HW233 slider loose on the plate. The finger contact is at the end of the arm and does not show from this angle.

19 July

Spent the day finding out that “force matching” names at least four unrelated paradigms. Only one of them measures sensory attenuation, and it is the one that needs a torque motor, a target force delivered to the finger, and two matching conditions. The others share nothing but the phrase.

This is worth a day because the failure mode is quiet. You can build a rig correctly to the wrong specification, run it, and get clean data that answers a question you were not asking. So the lineage went into its own document, with the excluded namesakes listed explicitly, before any metal was cut.

20 July

Rewrote the calibration tool. The version I had asked me to type in the readings, which meant the numbers that decided whether a calibration passed were transcribed by the person who wanted it to pass.

The new one measures eight criteria: gain, offset, R², RMSE, maximum absolute error, hysteresis, tare shift and drift. It issues a calibration ID, evaluates each criterion against a threshold, and if any of them fail it declines to write the calibration file. That last part matters more than the arithmetic. A failed calibration that is merely reported can still be used by accident; one that was never written cannot.

22 July

Had to choose a lineage for the indirect condition, and the two available choices are not equivalent.

McNaughton maps the slider to motor position, linearised with a fourth-order polynomial. The older UCL work maps the indirect input to a force servo, so the motor is commanded to produce a force rather than to travel to a place. Both paths existed in my code because I had written both while working out which was which.

I locked the force servo as the primary mode. It puts the two conditions on the same physical quantity: the participant produces a force directly, or asks for a force through the slider, and the comparison is force against force rather than force against a position that stands in for one. The position path stays in the code, unused. It will be reported as untested rather than as rejected, because I never ran it against anything.

27 July

Found that my own lineage document was out of date, which is a specific embarrassment when the document exists to prevent exactly that.

The box at the top still described the rig as belonging to the McNaughton line. That had been true when I wrote it and stopped being true five days earlier when I locked the force servo. The rig takes McNaughton’s timing, target forces and analysis window, and the UCL topology for how the indirect condition delivers force. It is a hybrid, and calling it a replication of either one is wrong in both directions.

Corrected the box and added the date, so the correction is visible rather than silently folded in.

The same day, a measurement problem in the slider itself. I had been modelling the relationship between the potentiometer’s output and its physical travel as a straight line. It is not one, and the straight line was leaving 13.2 per cent deviation.

That sits in exactly the wrong place. It is the step that turns slider position into a force target, so it sits between what the participant asks for and what the motor delivers, in the condition whose whole purpose is to separate those two. Replaced it with piecewise-linear interpolation. The old linear coefficients are still written to the record but no longer compute anything, because a transformation that has been shown wrong once should not be able to come back quietly.

28 July

Closed the calibration. Traced the force channel against dead weights over the whole range the task uses, 0 to 4.066 N, which puts every experimental target inside measured territory with nothing left to extrapolation. The fit gives R² = 0.9998 and an RMSE of 0.021 N, against a threshold of 0.05.

The number I did not expect was hysteresis. I had measured rising and falling loads as separate sequences instead of pooling them, on the argument that direction-dependent error is real and pooling is how you hide it from yourself. It came to 0.0498 N against a 0.05 N ceiling. It passes by 0.24 mN. Had I pooled the sequences the way the older tool did, that would have vanished into the RMSE and I would have recorded a comfortably clean run.

Seven criteria measured, seven passed. The eighth is drift, and I could not measure it at all: drift is a comparison between two calibrations and this is the first one. So the honest statement is not that the instrument passed eight checks. It is that it passed the seven that a single session can produce, and that this record is now the reference the next session gets compared against.

Also settled the contact pad. Earlier runs had no silicone pad, then had one, then briefly did not again. Since the pad changes what the finger feels, none of that engineering data can go in the analysis pool, and the confirmatory collection runs on the fixed configuration with the pad. That decision costs nothing now and would have cost the study later.